OPERATION MANUAL DAKOTA ULTRASONICS DDFFXX--77 Ultrasonic Flaw Detector (Manual 2 of 2) P/N P-220-0002 Rev

Strona 2

Dakota Ultrasonics 6 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DEFAULT SETUP is highlighted. 3) Press the ENTER key

Strona 3

Dakota Ultrasonics 96 Grid File Formats Sequential Log Formats Important Note: For the duration of this chapter, all reference

Strona 4

DFX-7 Ultrasonic Flaw Detector 97 Creating a Name Grid/Seq Log Name : Can contain a combination of up to 20 numeric, alpha, or special characters lis

Strona 5

Dakota Ultrasonics 98 Grid Log Sequential Log 4) Press the ENTER key to display the new Grid or Seq Edit Box. 5) Use

Strona 6 - 1.2 Electrical Warning

DFX-7 Ultrasonic Flaw Detector 99 9) Use the CLR key to backspace if necessary. 10) Repeat steps 6 - 9 until the Grid or Seq Name (File Name) is com

Strona 7 - 2.1 Gauge Type

Dakota Ultrasonics 100 3) Use the UP, DOWN, LEFT, & RIGHT arrow keys to highlight the appropriate alpha characters. 4) Press the ENTER key to sel

Strona 8

DFX-7 Ultrasonic Flaw Detector 101 Sequential: The sequential file format can be viewed as a file as a single column of up to 512 possible rows (read

Strona 9 - Load Default Setup

Dakota Ultrasonics 102 Grid Log Sequential Log 3) GRID LOG: Use the LEFT, & RIGHT arrow keys to scroll the Colum

Strona 10

DFX-7 Ultrasonic Flaw Detector 103 Grid Log Sequential Log 1) Use the UP and DOWN arrow keys to scroll through the new

Strona 11 - CHAPTER THREE

Dakota Ultrasonics 104 and advance to the next character field, in conjunction with using the CLR key to backspace if necessary. 4) Press the OK key t

Strona 12

DFX-7 Ultrasonic Flaw Detector 105 Selecting the Auto Increment Direction The Auto Increment feature gives the user the ability to specify which dire

Strona 13 - 3.3 CAL – Menu

7 CHAPTER THREE KEYBOARD, MENU, & CONNECTOR REFERENCE 3.1 Menu Key (Operation & Sub Menus) The Menu key activates the primary menu

Strona 14 - 3.5 TUNE – Menu

Dakota Ultrasonics 106 Saving Graphics The DFX-7 provides the user with the ability to save a snapshot of the display screen and all the current setti

Strona 15 - 3.7 GT2 – Menu

DFX-7 Ultrasonic Flaw Detector 107 Saving a Grid Once all the parameters are set, the user has the option of saving or canceling the new grid.

Strona 16 - 3.8 TRIG – Menu

Dakota Ultrasonics 108 Grid Log Sequential Log 6) Press the OK key to save the New Grid or Seq Log, or the ESC key to cancel the New

Strona 17 - 3.11 DAC – Menu

DFX-7 Ultrasonic Flaw Detector 109 Grid Log Sequential Log 1) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll

Strona 18 - 3.13 SETUP – Menu

Dakota Ultrasonics 110 Viewing Stored Waveforms Grid Log Sequential Log 1) Press the MEAS key once to activate measu

Strona 19 - 3.14 DATA – Menu

DFX-7 Ultrasonic Flaw Detector 111 5) Abort the Grid/Seq Log View Box by pressing the MEAS key at any time. 11.5 Deleting Grids (Files) Deleting One

Strona 20 - 3.19 OK Key

Dakota Ultrasonics 112 3) Press the ENTER key to display the File List Box. 4) Use the UP and DOWN arrow keys to scroll through the stored Files un

Strona 21

DFX-7 Ultrasonic Flaw Detector 113 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DELETE ALL DATA is highlighted. 3) P

Strona 22

Dakota Ultrasonics 114 11.6 Editing a Grid (File) Once a grid has been created and saved to memory, the user can edit the Comments or Increment Direct

Strona 23 - 3.26 Navigating the Hot Menu

DFX-7 Ultrasonic Flaw Detector 115 5) Press the ENTER key to activate the Alpha Edit box – Only used when editing the NOTE. 6) Use the UP, DOWN, LEFT

Strona 24 - Hot Menu

Dakota Ultrasonics 8 Activating and Navigation of the Menu Items 1) Press the MENU key once to activate the menu items tab. Press the MENU key multi

Strona 25

Dakota Ultrasonics 116 10) Press the UP or DOWN arrow key to highlight SAVE CHANGES, and the OK key to activate the confirmation screen. 11) Press

Strona 26

DFX-7 Ultrasonic Flaw Detector 117 Grid Log Sequential Log 1) Press the MENU key once to activate the menu items tab. Press the MENU

Strona 27 - Display View

Dakota Ultrasonics 118 5) Press the ENTER key to activate the confirmation screen. 6) Press the OK key to load the file from memory. 7) Press the MEA

Strona 28 - Setting Delay

DFX-7 Ultrasonic Flaw Detector 119 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until CLOSE is highlighted. 3) Press t

Strona 29 - Setting Range

120 CHAPTER TWELVE SETUPS – CREATE, STORE, EDIT, & RECALL 12.1 Introduction to Setups Often times, users are faced with a variety of tasks and app

Strona 30

DFX-7 Ultrasonic Flaw Detector 121 Opening a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to

Strona 31 - Adjusting Gain

Dakota Ultrasonics 122 5) Press the ENTER key to activate the confirmation screen. 6) Press the OK key to load the setup from memory. 7) Press the ME

Strona 32

DFX-7 Ultrasonic Flaw Detector 123 Saving a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to ta

Strona 33

Dakota Ultrasonics 124 5) When the parameter to edit is highlighted, press the ENTER key to activate the Alpha Edit Box. 6) Use the UP

Strona 34 - 4.4 Gates

DFX-7 Ultrasonic Flaw Detector 125 11) Use the UP and DOWN arrow keys to scroll through the setups until the target location to save the Setup is h

Strona 35 - Activating a Gate

DFX-7 Ultrasonic Flaw Detector 9 3.2 Probe – Menu TYPE: Enables the user to select the type of transducer being used, either Single or Dual. PULSE

Strona 36

Dakota Ultrasonics 126 Deleting a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right a

Strona 37

DFX-7 Ultrasonic Flaw Detector 127 5) When the Setup Name is highlighted, press the ENTER key to display the confirmation screen. 6) Press the OK key

Strona 38 - Dakota Ultrasonics

Dakota Ultrasonics 128 Using the Default Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab r

Strona 39 - DFX-7 Setup

DFX-7 Ultrasonic Flaw Detector 129 12.6 Selecting a Language The DFX-7 is equipped with a language option. Currently, the only languages supported ar

Strona 40

130 CHAPTER THIRTEEN USING THE UTILITY SOFTWARE 13.1 Computer System Requirements DakView will run on many different operating systems: Windows XP, W

Strona 41

DFX-7 Ultrasonic Flaw Detector 131 13.3 Using the XFER menu (DFX-7) The XFER menu of the DFX-7 will be used in conjunction with the DakView PC softwa

Strona 42

132 APPENDIX A - VELOCITY TABLE Material L-Wave velocity in/µs L-Wave velocity m/s S-Wave Velocity in/µs S-Wave Velocity in/µs Aluminum 0.2510 63

Strona 43

DFX-7 Ultrasonic Flaw Detector 133 Steel Stainless" 0.2230 5664 .1230 3120 Teflon 0.0540 1372 .2500 6350 Tin 0.1310 3327 .0660 1670 Titanium 0.

Strona 44

134 APPENDIX B - SETUP LIBRARY Num Name Comment 1 Gn/AGC Velocity 1 Enter Custom Name 2 … 3 … 4 … 5 … 6 … … …

Strona 45

135 WARRANTY INFORMATION • Warranty Statement • Dakota Ultrasonics warrants the DFX-7 against defects in materials and workmanship for a period of t

Strona 46

Dakota Ultrasonics 10 3.4 DISP (display) – Menu VIEW: Selectable RF, +RECT, -RECT, RECTFW views. Refer to page 23 for further info. DELAY: Pro

Strona 47

DFX-7 Ultrasonic Flaw Detector 11 be considered as similar to turning the volume up or down on a stereo receiver. Refer to page 27 for further info.

Strona 48

Dakota Ultrasonics 12 ALARM: Toggles the alarm function of gate 2; on, off, audible. The audible feature sounds a beeper when the waveform breaks th

Strona 49

DFX-7 Ultrasonic Flaw Detector 13 3.9 AWS – Menu AWS: This feature toggles the AWS feature on/off. Used in the analysis of welds according to Am

Strona 50 - 5.3 Angle Beam Calibration

Dakota Ultrasonics 14 DRAW: Toggles the DRAW mode on/off/audible, and used to draw the DAC curve. Note: TCG curve cannot be drawn in RF view. Refe

Strona 51

DFX-7 Ultrasonic Flaw Detector 15 LANGUAGE: Provides the user the ability to select different languages for the DFX-7. Refer to page 129 for further

Strona 53

Dakota Ultrasonics 16 BACKUP SETUPS: Enables the user the ability to backup the setups currently stored in the DFX-7 to a PC via RS232 port. USB ad

Strona 54 - 6.2 Additional Comments

DFX-7 Ultrasonic Flaw Detector 17 3.20 ESC Key The ESC key is used in the MENU, MEAS, and EDIT functions as a back or escape function. If the DFX-

Strona 55 - 6.3 Setup Trigonometry Mode

Dakota Ultrasonics 18 In order to understand how to operate the DFX-7, it’s best to start off with an understanding of what we’re looking at. The D

Strona 56

DFX-7 Ultrasonic Flaw Detector 19 • Detect i. Zero crossing (ZX) ii. Flank (FK) iii. Peak (PK) • Leg – Skip leg reference, (L1…L4…) • High Speed Sc

Strona 57

Dakota Ultrasonics 20 purpose of the design, was to provide the user with an efficient way to make adjustments on the fly during inspections. The fol

Strona 58

DFX-7 Ultrasonic Flaw Detector 21 6) Repeat steps 4 & 5 until the start position is correctly displayed. 7) Press the OK key to set the WIDTH1 pos

Strona 59 - 7.3 Setup AWS

Dakota Ultrasonics 22 pin serial cable. Note: This connector is also used to upgrade the DFX-7 with the latest version of firmware. USB Serial to US

Strona 60 - Activating & Setting AWS

23 CHAPTER FOUR SETTING UP FOR MEASUREMENT 4.1 Selecting Display View The DFX-7 includes a variety of viewing options for the waveform display. There

Strona 61

Dakota Ultrasonics 24 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until VIEW is highlighted. 3) Use the LEFT and RIGHT arro

Strona 62

DFX-7 Ultrasonic Flaw Detector 25 Setting Delay 1) Press the MEAS key once to activate measure menu items. Press the MEAS key multiple times to mov

Strona 63

CHAPTER ONE INTRODUCTION ... 1CHAPTER TWO QUICK START GUIDE ...

Strona 64 - 8.2 Additional Comments

Dakota Ultrasonics 26 The Range is the distance from the delay that will be viewable on the screen. It can be adjusted to any value greater than the

Strona 65 - 8.3 Creating a DAC Curve

DFX-7 Ultrasonic Flaw Detector 27 4.3 Gain The gain feature of the DFX-7 is used to adjust the signal amplitude to a predetermined reference height

Strona 66

Dakota Ultrasonics 28 2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted value. 3) Alternatively, press the ENTER key to dis

Strona 67 - Drawing the DAC

DFX-7 Ultrasonic Flaw Detector 29 The user can also access and adjust the gain from the tabbed menus. However, this method is more tedious than makin

Strona 68

Dakota Ultrasonics 30 9) Finally, press the MEAS key to return to the measurement screen. 4.4 Gates The gates are an important part of any ultrason

Strona 69

DFX-7 Ultrasonic Flaw Detector 31 No Gate Gate On The screenshots above show a rectified waveform with and without a gate enabled. The following st

Strona 70

Dakota Ultrasonics 32 Note: Above will trigger the alarm when a signal meets or exceeds the height of the gate, and below will detect on anything tha

Strona 71

DFX-7 Ultrasonic Flaw Detector 33 6) Press the OK key to set the GATE1 WIDTH and return to the menu screen, or ESC to cancel entering the GATE1 WIDTH.

Strona 72 - Drawing the TCG

Dakota Ultrasonics 34 Note: Sets where the detection will occur, on the negative or positive going echoes.

Strona 73

35 CHAPTER FIVE CALIBRATION In this section we’ll focus on calibrating the DFX-7. There are a number of ways to go about performing a calibration, bu

Strona 75 - Adjusting Velocity

Dakota Ultrasonics 36 The gauge will require a few basic adjustments in order to get underway with the calibration process. For this example, in o

Strona 76 - Adjusting Zero

DFX-7 Ultrasonic Flaw Detector 37 2) Press the ENTER key to display the Digits Edit Box. 3) Press the UP and DOWN arrow keys to scroll the highlighte

Strona 77

Dakota Ultrasonics 38 Activating a Gate 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right an

Strona 78

DFX-7 Ultrasonic Flaw Detector 39 5) Repeat steps 3 & 4 until the desired GATE1 value is displayed. Note: In our example, a value less than 1 inc

Strona 79 - Using Auto Find

Dakota Ultrasonics 40 Setting Polarity 1) Use the UP and DOWN arrow keys to scroll through the sub menu items until POLARITY1 is highlighted. 2) Press

Strona 80 - Adjusting Brightness

DFX-7 Ultrasonic Flaw Detector 41 Coupling to the 1” Cal Standard (1st Point) Apply a drop of couplant on the transducer and place the transducer in s

Strona 81 - Selecting a Color Scheme

Dakota Ultrasonics 42 3) Press the ENTER key to display the Digits Edit Box. 4) Press the UP and DOWN arrow keys to scroll the highlighted value. 5)

Strona 82

DFX-7 Ultrasonic Flaw Detector 43 stable. Also, be sure there is sufficient signal amplitude in order to avoid ‘peak jumping’ or missing the first cy

Strona 83 - 10.7 DIM

Dakota Ultrasonics 44 4) Press the UP and DOWN arrow keys to scroll the highlighted value. 5) Press the LEFT and RIGHT arrow keys to scroll the digit

Strona 84

DFX-7 Ultrasonic Flaw Detector 45 Adjusting the Gate to Calibrated Range 1) Press the MEAS key once to activate the measure menu items. Press the ME

Strona 85 - Setting the Look & Feel

CHAPTER ONE INTRODUCTION The Dakota Ultrasonics model DFX-7 is both, an ultrasonic flaw detector, as well as an ultrasonic A/B Scan thickness scope

Strona 86 - 10.10 Polarity

Dakota Ultrasonics 46 Note: CHECK YOUR CALIBRATION! Place the transducer back on the one inch point, and any others in the given range, to confirm l

Strona 87 - Toggle Polarity (+/-)

DFX-7 Ultrasonic Flaw Detector 47 Shear versus Longitudinal Velocity: Unlike the previous straight beam calibration, which uses a longitudinal wave, a

Strona 88

Dakota Ultrasonics 48 wave values, the TRIG mode must be enabled. The following outlines the steps to enable the shear wave velocity table and select

Strona 89 - Selecting the Detect

DFX-7 Ultrasonic Flaw Detector 49 7) Press the ENTER key to select the material type and initiate the overwrite confirmation screen. 8) Press the OK k

Strona 90

50 CHAPTER SIX TRIGONOMETRY MODE 6.1 Introduction to TRIG Trig mode is most commonly used for inspecting welded joints using an angle beam transducer.

Strona 91

DFX-7 Ultrasonic Flaw Detector 51 When TRIG mode is enabled, the calculated measurements will be displayed below the waveform display area at (A), an

Strona 92

Dakota Ultrasonics 52 Activating & Setting the Parameters Activate Trig Mode 1) Press the MENU key once to activate the menu items tab. Press

Strona 93 - Setting the Max PRF

DFX-7 Ultrasonic Flaw Detector 53 4) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 5) Repeat steps 3 & 4 until the thickn

Strona 94

Dakota Ultrasonics 54 6) Press the OK key to set the X-OFFSET and return to the menu screen, or ESC to cancel entering the X-OFFSET. Note: The X-Offs

Strona 95

55 CHAPTER SEVEN AWS – WELD INSPECTION 7.1 Introduction to AWS AWS provides a toolkit for evaluating defects according to the American Welding Society

Strona 96

1.2 Electrical Warning The DFX-7 contains a high voltage pulser. It’s recommended that the gauge be powered off prior to connecting or disconnecting

Strona 97

Dakota Ultrasonics 56 inspection, and should additionally be setup prior to enabling the AWS feature. Refer to Chapter Six for instructions on settin

Strona 98

DFX-7 Ultrasonic Flaw Detector 57 3) Press the UP and DOWN arrow keys to scroll the highlighted value. 4) Press the LEFT and RIGHT arrow keys to scrol

Strona 99 - CHAPTER ELEVEN

Dakota Ultrasonics 58 2) Press the LEFT and RIGHT arrow keys to position the start of GATE1 to trigger on the waveform. 3) Optionally, press the MEAS

Strona 100

DFX-7 Ultrasonic Flaw Detector 59 Note: The AWS screen can be toggle on by pressing the OK key while in the main measurement screen. Equally, the Ho

Strona 101 - Creating a Name

60 CHAPTER EIGHT DISTANCE AMPLITUDE CORRECTION (DAC) 8.1 Introduction to DAC A DAC curve is a gate with a variable threshold to compensate for a given

Strona 102

DFX-7 Ultrasonic Flaw Detector 61 %FSH – The top of the waveform display represents 100% full screen height. Any waveform amplitudes generated greate

Strona 103 -

Dakota Ultrasonics 62 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until DAC is highlighted. 3) Press the LEFT and RIGHT arr

Strona 104

DFX-7 Ultrasonic Flaw Detector 63 2) Proceed to the next section – Drawing the DAC. Drawing the DAC 1) With all the parameters set in the prev

Strona 105

Dakota Ultrasonics 64 2) Press the MEAS key multiple times to move right, and the ESC key multiple times to move left until the DAC BOX: cell is highl

Strona 106

DFX-7 Ultrasonic Flaw Detector 65 7) Press the UP and DOWN arrow keys until DRAW is highlighted, followed by pressing the LEFT and RIGHT arrow keys u

Strona 107

3 CHAPTER TWO QUICK START GUIDE This section will cover a basic quick start guide to initially get up and running with the DFX-7 for demonstration pur

Strona 108

66 CHAPTER NINE TIME CORRECTED GAIN (TCG) 9.1 Introduction to TCG The TCG feature increases gain as distance increases, in order to achieve an overall

Strona 109

DFX-7 Ultrasonic Flaw Detector 67 Activating TCG 4) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab

Strona 110 - Saving Graphics

Dakota Ultrasonics 68 11) Press the OK key to set the % SCRN HEIGHT and return to the menu screen, or ESC to cancel entering the % SCRN HEIGHT. Acti

Strona 111 - Saving a Grid

DFX-7 Ultrasonic Flaw Detector 69 ex. 1 ex. 2 2) Press the MEAS key multiple times to move right, and the ESC key multiple t

Strona 112 - 11.3 Storing a waveform

Dakota Ultrasonics 70 7) Press the UP and DOWN arrow keys until DRAW is highlighted, followed by pressing the LEFT and RIGHT arrow keys until OFF is

Strona 113 - DFX-7 will automatically

71 CHAPTER TEN ADDITIONAL FEATURES OF THE DFX-7 10.1 Material Velocity The DFX-7 has a built in feature to manually adjust the material velocity to a

Strona 114 - Viewing Stored Waveforms

Dakota Ultrasonics 72 6) Press the LEFT and RIGHT arrow keys to scroll the digit locations. 7) Repeat steps 5 & 6 until the velocity value is co

Strona 115 - Deleting One File

DFX-7 Ultrasonic Flaw Detector 73 2) Use the UP and DOWN arrow keys to scroll through the sub menu items until ZERO is highlighted. 3) Press the LEFT

Strona 116 - Deleting All Data

Dakota Ultrasonics 74 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mult

Strona 117

DFX-7 Ultrasonic Flaw Detector 75 6) Finally, press the MEAS key to return to the measurement screen and begin taking readings. 10.4 Auto Find The D

Strona 118 - Editing a Grid

Dakota Ultrasonics 4 Pulse: Thin Matl 1pt: Uncal Delay: 0.000 Gain: 65 Gate1: 0.484 Max PRF: 66 Matl 2pt: Uncal Range: 5.011 Detect Mode

Strona 119

Dakota Ultrasonics 76 4) Press the ENTER key to automatically adjust the display settings and bring the waveform into view. 5) Once the signal is dis

Strona 120

DFX-7 Ultrasonic Flaw Detector 77 3) Press the LEFT and RIGHT arrow keys to scroll the value. When the correct brightness value is being displayed, p

Strona 121

Dakota Ultrasonics 78 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mult

Strona 122

DFX-7 Ultrasonic Flaw Detector 79 6) Finally, press the MEAS key to return to the measurement screen and begin taking readings. 10.7 DIM The DFX-7 h

Strona 123

Dakota Ultrasonics 80 4) Finally, press the MEAS key to return to the measurement screen and begin taking readings. 10.8 Graphics Options (look &

Strona 124 - 12.2 Opening a Setup

DFX-7 Ultrasonic Flaw Detector 81 The procedure to select the look and feel option is outlined below: Setting the Look & Feel 1) Press th

Strona 125

Dakota Ultrasonics 82 10.9 Detect Mark Another look and feel option of the DFX-7 is the DETECT MARK feature. The detect mark give the user the option

Strona 126 - 12.3 Saving a Setup

DFX-7 Ultrasonic Flaw Detector 83 The diagram above is an RF view, showing both the positive and negative going cycles for purposes of explanation.

Strona 127 - Saving a Setup

Dakota Ultrasonics 84 Note: The screenshots above show where all the possible locations for polarity selection are found in the DFX-7. For more i

Strona 128

DFX-7 Ultrasonic Flaw Detector 85 Z-Cross Flank Peak Refer to the diagrams above. The current polarity is positive, as shown by the position of t

Strona 129 - 12.4 Deleting a Saved Setup

DFX-7 Ultrasonic Flaw Detector 5 FILE % SCRN HEIGHT CURVE REFLECT LOSS DEFAULT SETUP CLOSE RESTORE DATA TRIGGER MATERIAL LOSS LANGUAGE DELETE ONE FI

Strona 130 - Deleting a Setup

Dakota Ultrasonics 86 7) Use the LEFT and RIGHT arrow keys to scroll the DETECT options. 8) Once the DETECT is displayed, press the MEAS key to return

Strona 131 - 12.5 Using the Default Setup

DFX-7 Ultrasonic Flaw Detector 87 4) Once the Pulse is displayed, press the MEAS key to return to the measurement screen. 10.13 Pulser Voltage The DF

Strona 132 - Using the Default Setup

Dakota Ultrasonics 88 4) Once the desired Pulser Voltage is displayed, press the MEAS key to return to the measurement screen. 10.14 Screen Update (

Strona 133

DFX-7 Ultrasonic Flaw Detector 89 10.15 Pulse Repetition Frequency (PRF) The DFX-7 is equipped an adjustable pulse repetition rate, with adjustable s

Strona 134 - 13.2 Installing DakView

Dakota Ultrasonics 90 8) Once the desired PRF is displayed, press the MEAS key to return to the measurement screen. 10.16 Damping The DFX-7 has an a

Strona 135

DFX-7 Ultrasonic Flaw Detector 91 4) Once the desired DAMPING is displayed, press the MEAS key to return to the measurement screen. 10.17 Filters (W

Strona 136 - VELOCITY TABLE

Dakota Ultrasonics 92 4) Once the desired FILTERS is displayed, press the MEAS key to return to the measurement screen. 10.18 Peak Hold The DFX-7 in

Strona 137

DFX-7 Ultrasonic Flaw Detector 93 1) Press the CLR key at any time to clear the currently displayed waveform, and continue the scanning process. Disab

Strona 138 - SETUP LIBRARY

Dakota Ultrasonics 94 Note: Assuming the operator currently has a waveform that they’d like to freeze on the display. 1) Press the MULTI MODE key loc

Strona 139 - WARRANTY INFORMATION

95 CHAPTER ELEVEN DATA STORAGE – SETUP, EDIT, & VIEW FILES 11.1 Introduction to Grid and Sequential file formats The DFX-7 is equipped with two da